• ORP Online Meter for RO Water Treatment, Model: pH ORP-5500
  • ORP Online Meter for RO Water Treatment, Model: pH ORP-5500
  • ORP Online Meter for RO Water Treatment, Model: pH ORP-5500

ORP Online Meter for RO Water Treatment, Model: pH ORP-5500

Display: Digital
Installation: Direct Connected
Usage: Multi-Functional Energy Meter
Electric Equipment: Single-phase
Theory: Electronic Meter
Power Meter Connection: Terminal
Customization:
Gold Member Since 2015

Suppliers with verified business licenses

Trading Company

Basic Info.

Model NO.
5500
Measured Signal
Continuous Wave
Type
Power Meter
Certification
RoHS, ISO, CE
Customized
Non-Customized
Voltage
AC 110V/AC220V/DC24V
pH Rang
0-14 Ppm
Orp Rang
-2000mv~2000mv
Temp.
(0.0~50.0)&Ordm
Transport Package
Carton Box
Specification
8 pcs per box
Trademark
Create
Origin
China
HS Code
8421
Production Capacity
50000 PCS Per Month

Product Description

It specializes in the fields of water purification inline monitoring, process water integration management system, environmental protection and industry wastewater treatment process management systems.

Technical Features
 
Model pH/ORP-5500 series pH/ORP online transmitting controller
Measurement range pH 0.00~14.00
ORP -2000mV~2000mV
Temp. (0.0~50.0)ºC(temperature compensation component:NTC10K)
Resolution pH 0.01
ORP 1mV
Temp. 0.1ºC
Accuracy pH 0.1
ORP ±5mV(electronic unit)
Temp. ±0.5ºC
Approximate input impedance 3×1011Ω
Buffer solution pH value: 10.00;9.18;7.00;6.86;4.01;4.00
Temp.compensation range (0~50)ºC(with 25ºC as standard) 
Manual and automatic temperature compensation
(4~20)mA Characteristics Isolated,fully adjustable,reverible,instrument/transmitter for selection
Loop resistance 500Ω(Max),DC 24V
Accuracy ±0.1mA
Control contact Electrical contacts Double relay SPST-NO,return model
Loop capacity AC 220V/AC 110V 2A(Max);DC 24V 2A(Max)
Power consumption <3W
Working environment temperature (0~50)ºC
humidity ≤85%RH(none condensation)
Storage environment Temp.(-20-60) ºC;relative humidity:≤85%RH(none condensation
Outline dimension 96mm×96mm×105mm(H×W×D)
Hole dimension 91mm×91mm(H×W)
Installation Panel mounted,fast installation
 
 
Order direction :
 
Basic order units :
pH/ORP-5500 pH/ORP Meter (DC24V)
pH/ORP-5500 pH/ORP Meter (AC110V)
pH/ORP-5500 pH/ORP Meter (AC220V)


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now